電感耦合等離子體光譜儀的保養(yǎng):

The maintenance of inductively coupled plasma spectrometer:

(1)氣體控制系統(tǒng):ICP的氣體控制系統(tǒng)是否穩(wěn)定正常地運(yùn)行,直接影響到儀器測(cè)定數(shù)據(jù)的好壞,如果氣路中有水珠或其它固體雜質(zhì)等都會(huì)造成氣流不穩(wěn)定,因此,對(duì)氣體控制系統(tǒng)要經(jīng)常進(jìn)行檢查和維護(hù)。首先要做氣體密封試驗(yàn),開(kāi)啟氣瓶及減壓閥,使氣體壓力指示在額定值上,然后關(guān)閉氣瓶,觀察減壓閥上的壓力表指針,應(yīng)在幾個(gè)小時(shí)內(nèi)沒(méi)有下降或下降很少,否則說(shuō)明氣路中有漏氣現(xiàn)象,需要檢查和排除。

(1) the gas control system: ICP gas control system is stable in normal operation, directly affect the determination of the quality of the data, if there is water or other solid impurities in the gas path will cause the flow instability, therefore, the gas control system to carry out regular inspection and maintenance. The first thing to do gas seal test, open the gas cylinder and the pressure relief valve, the gas pressure indicator in the ratings, and then close the cylinder, observe the valve on the pressure gauge pointer should not drop or fall, rarely in a few hours, otherwise the leakage phenomenon in the gas path, need to check and troubleshooting.

(2)使用環(huán)境:電感耦合等離子體光譜儀與其它大型精密儀器一樣,需要在一定的環(huán)境條件下運(yùn)行,否則,不僅影響儀器的性能,甚至造成損壞,縮短壽命等。根據(jù)光學(xué)儀器的特點(diǎn),對(duì)環(huán)境溫度和濕度有一定要求。如果溫度變化太大,光學(xué)組件受溫度變化的影響就會(huì)產(chǎn)生譜線漂移,造成測(cè)定數(shù)據(jù)不穩(wěn)定;而如果環(huán)境濕度過(guò)大,儀器的光學(xué)部件,特別是光柵容易受潮損壞或性能降低

(2) the use of the environment: inductively coupled plasma atomic emission spectrometer and other large-scale precision instruments, need to run under certain environmental conditions, otherwise, not only affects the performance of the instrument, or even cause damage, shorten the life. According to the characteristics of optical instruments, the environmental temperature and humidity have certain requirements. If the temperature change is too large, the impact will be affected by temperature change of the optical components of spectral line shift caused by measured data instability; and if the humidity is too large, optics instrument, especially grating easy to damp damage or reduced performance

(3)每月用超聲波清洗霧化室、炬管、樣品噴射管;

(3) a month with ultrasonic cleaning spray chamber and the torch tube, sample injection pipe;

(4)每四個(gè)月更換光譜儀上的風(fēng)扇過(guò)濾網(wǎng);

(4) fan filter on the spectrometer for every four months;

(5)測(cè)試有機(jī)樣品后用軟布清潔炬室,防止有機(jī)成分沉積;

(5) test of organic samples with a soft cloth to clean the torch room, to prevent the organic components of the sediment;

(6)進(jìn)樣管、蠕動(dòng)泵管、廢液管、墊圈等消耗品根據(jù)使用情況及時(shí)更換;

(6) sampling tube and peristaltic pump tube, waste pipe, gasket and other consumables according to the use of timely replacement;

(7)每年根據(jù)儀器公司提供的日常維護(hù)條款對(duì)儀器做全面保養(yǎng)。

(7) in accordance with the provisions of the daily maintenance of the instrument company to do a comprehensive maintenance of the instrument.

所以使用電感耦合等離子體光譜儀一定要好好保護(hù)儀器,要合理的進(jìn)行保養(yǎng)與注意,出現(xiàn)故障就不要用了,有問(wèn)題就要進(jìn)行修理。

So the use of inductively coupled plasma spectrometer must be a good protective equipment, to a reasonable maintenance and attention, there is no need to use the fault, there is a problem to be repaired.

電感耦合等離子體光譜儀注意事項(xiàng): 送樣要求:液體,來(lái)樣標(biāo)明名稱(chēng)、主要成分、待測(cè)元素大致含量。

Inductively coupled plasma spectrometer matters needing attention: sample delivery requirements: liquid, to indicate the name of the sample, the main component, the approximate content of the element to be measured.

樣品要求:

Sample requirement:

(1)試液需完全澄清。

(1) solution to fully clarify.

(2)試液需pH≤7,鹽酸或硝酸濃度≤10%。

(2) pH was less than 7, hydrochloric acid or nitric acid concentration is less than 10%.

(3)試液中基體元素含量不得超過(guò)待測(cè)元素一萬(wàn)倍。

(3) the content of matrix in the test shall not exceed ten thousand times the measured elements.

(4)試液中若含有有機(jī)物必須作硝化處理后上機(jī)測(cè)定。

(4) if the solution containing organic matter determination must operate nitrification.

(5)樣品量:根據(jù)儀器要求

(5) sample size: according to the requirements of the instrument