1 開(kāi)機(jī)順序:實(shí)驗(yàn)室總電源→氬氣→冷卻水→抽風(fēng)機(jī)→儀器主機(jī)→進(jìn)樣泵→電腦→smart analyzer vision工作站

1 boot order: total power, argon, laboratory, exhaust fan, cooling water, injection pump, instrument host computer, smart analyzer vision workstation

2 方法編輯:點(diǎn)擊smart analyzer vision工作站左邊工具箱中的method按鈕進(jìn)入方法編輯窗口,點(diǎn)擊new method,輸入方法名,ok↓確定。

2 methods of editing: click on the smart vision workstation analyzer left in the toolbox method button to enter the method of editing window, click on the new method, the input method name, OK: sure.

進(jìn)行測(cè)量

To measure

當(dāng)方法編輯好后,就可以進(jìn)行點(diǎn)火開(kāi)始測(cè)量。點(diǎn)火:點(diǎn)擊system/setup devices,分別設(shè)置等離子氣體流量為10l/min、輔助氣流量為1l/min、霧化器氣流量為1l/min,點(diǎn)擊apply 進(jìn)行沖洗氣路2分鐘。在矩管圖中觀察以上各氣流量回零后,點(diǎn)擊plasma點(diǎn)燃等離子體,點(diǎn)火后點(diǎn)擊pump啟動(dòng)進(jìn)樣泵。在進(jìn)樣前吸入蒸餾水或空基體溶液5∽15分鐘,以加熱和平衡組成成分。

When the method is edited, the ignition can be started. Ignition: click on the system/setup devices, were set up plasma gas flow rate of 10l/min, auxiliary gas flow 1l/min, nebulizer gas flow rate of 1l/min, click apply to wash gas for 2 minutes. In the torque tube chart above the gas flow back to zero, click plasma to ignite the plasma, ignition after the click pump to start the sample pump. In the sample before inhalation of distilled water or 5 ~ 15 minutes free matrix solution, composition for heating and balance.

4、數(shù)據(jù)處理

4, data processing

5、關(guān)機(jī),順序關(guān)機(jī):關(guān)氬氣→冷卻循環(huán)水→儀器主機(jī)→抽風(fēng)機(jī)→電腦→實(shí)驗(yàn)室總電源。

5, shutdown, shutdown order: argon, cooling water, exhaust fan, computer, instrument, laboratory of general power.